ImmunoStrip® for CP4 EPSPS (RR)
This Agdia ImmunoStrip® for Roundup Ready® is designed and validated for detecting the presence or absence of the CP4 EPSPS (RR) protein in single seed and leaves and for testing bulk mixtures. The Agdia Roundup Ready® ImmunoStrip® is validated for testing transgenic corn, cotton, and soybean with a detection limit of 1 in 1000 seeds (0.1%).
These strips are commonly used by plant breeders, universities and private testing labs who want fast, clear results.
Agdia ImmunoStrip® for Roundup Ready® has been evaluated by the USDA, AMS, Federal Grain Inspection Service (FGIS) and was found to meet all test performance criteria as defined in the “Design Criteria and Test Performance Specifications for Biotechnology Rapid Test Kits,” May 2017 version. For further information see the FGIS certificate.
Roundup Ready® is a registered trademark of Bayer Group.
Click here for other product documents such as the User Guide or Validation Report.
Included in STX Item:
- ImmunoStrips®
- User Guide
Required but Not Included:
- SEB4 Extraction buffer
- Scissors
- Extraction tool
- Mesh sample bags, weigh paper, 48 west well microtiter plate, or seed crusher insert
- Letter holder or other device to hold sample extraction bags upright
For a full list of included items, including sizes, please see the Content List.
Validation Report Summary
Test Format: ImmunoStrip® Lateral Flow Device (LFD) |
Full Validation Report |
Diagnostic Sensitivity and Specificity |
---|
Total Samples: 432 Correct Diagnoses: 432 Number of Operators: 3 Number of Production Lots: 1 Percent: 100% |
Proteins Detected | |
---|---|
CP4 EPSPS |
Cross-reactions | |
---|---|
None Known |
For a list of validated hosts, please see the Full Validation Report.
Additional Documents | ||
---|---|---|