ImmunoStrip® for Bt-Cry1Ab/1Ac
Tech Note: Storage Condition Change
The Bt-Cry1Ab/1Ac ImmunoStrip® is designed to quickly determine the presence or absence of the Bt-Cry1Ab and Bt-Cry1Ac proteins expressed in transgenic corn and cotton. In just minutes, this test will detect the Bt-Cry1Ab protein expressed in corn events MON810 and Bt-11. This test is also appropriate for detecting Bt-Cry1Ac events in cotton. This Bt-Cry1Ab/Ac ImmunoStrip® test is appropriate for testing most transgenic plants expressing either Bt-Cry1Ab or Bt-Cry1Ac.
Click here for other product documents such as the User Guide or Validation Report.
Included in STX Item:
- ImmunoStrips®
- User Guide
- Tech Note
Required but Not Included:
- SEB4 Extraction buffer
- Scissors
- Extraction tool
- Mesh sample bags, weigh paper, 48 west well microtiter plate, or seed crusher insert
- Letter holder or other device to hold sample extraction bags upright
For a full list of included items, including sizes, please see the Content List.
Validation Report Summary
Full Validation Report
Test Format: ImmunoStrip® Lateral Flow Device (LFD) |
Diagnostic Sensitivity and Specificity |
---|
Total Samples: 2560 Correct Diagnoses: 2560 Number of Operators: 5 Number of Production Lots: 3 Percent: 100% |
Proteins Detected | |
---|---|
Bt-Cry1Ab | Bt-Cry1Ac |
Cross-reactions | |
---|---|
None Known |
For a list of validated hosts, please see the Full Validation Report.